Specifications
- Mag-01 & Mag-01H
|
| Electrical
|
 |
Mag-01
 |
Mag-01H
 |
Measurement
Range
Low Field Probes (B, C, &
F)
High Field Probes (D, E, &
G) |
0 to 200µT
0 to 2,000µT |
0 to 20µT / 200µT
0 to 200µT / 2,000µT |
Measurement
Resolution (Display)
Low Field Probes (B, C, &
F)
0 to 2µT
2 to 20µT
20 to 200µT
High Field Probes (D, E, &
G)
0 to 20µT
20 to 200µT
200 to 2,000µT
|
N/A
1nT
10nT
N/A
10nT
100nT
|
0.1nT
1nT
10nT
1nT
10nT
100nT
|
| Bandwidth |
0
to 10Hz, -12dB per octave roll off
|
| Calibration
Accuracy |
0.1%
|
| Resolution
(maximum) |
1nT
|
0.1nT
|
| Zero
Field Offset (at 20°C) |
<
±5nT
|
| Scaling
Temperature Coefficient |
<
10ppm/°C
|
| Offset
Temperature Coefficient |
<
0.01nT/°C
|
| Battery |
LC-R061R3PU,
lead acid, 6V, 1.3Ah
|
| Operating
Temperature |
-10°C
to +50°C
|
| Relative
humidity |
80%
non-condensing
|
|
Front Panel |
| On
/ Off Switch |
Switches
instrument on or off
|
| Probe
Input |
6
pole waterproof Fischer connector |
| Charger
Indicator |
Illuminated
when external supply connected
|
| Offset
Control |
N/A
|
Allows
offsets up to 90µT to be added to the reading
|
| Offset
Polarity Control |
N/A
|
Controls polarity of the offset addition
|
| Sensitivity
Control |
N/A
|
Increases the sensitivity by a factor of 10
|
| Rear
Panel |
| Battery
Charger Input |
2.1mm
co-axial socket, 6-18Vdc, 0.5A, polarity protected.
Continuous or intermittent use.
|
| Analog
Output |
4.0mm
insulated 'banana' sockets
|
Full
Scale Output Voltage
|
2.0
volts
|
Output
Impedance
|
1
k-ohm
|
Scaling
Low Field Probes (B, C, &
F)
High Field Probes (D, E, &
G) |
100µT/V
1,000µT/V
|
10µT/V / 100µT/V
100µT/V / 1,000µT/V
|
| Mechanical
|
| Enclosure |
High
impact ABS plastic |
| Dimensions |
155
x 170 x 68 |
| Weight
|
1.5kg
|
|
|
Specifications
- Mag Probes
|
| |
Low
Field
|
High
Field
|
| Range |
0
to 0.2mT
|
0
to 2.0mT
|
| Offset
Error |
<1nT
|
<20nT
|
|
Scaling Temperature Coefficient |
<±10ppm/°C
|
<±30ppm/°C
|
| Sensitive
Volume of Core |
0.02cm³
|
0.0036cm³
|
| Excitation
Power |
<26mW
|
<16mW
|
| Linearity |
<0.01%
|
| Calibration
Accuracy |
<1%
|
Probe
Alignment Error
(With reference to Case) |
<0.2°
|
| Operating
temperature |
| Axial
& Transverse |
-30°C
to +75°C
|
| Cryogenic |
Liquid
helium to +30°C
|
| Model
Numbers |
| Axial
|
Mag-B
|
Mag-D
|
| Transverse |
Mag-C
|
Mag-E
|
| Cryogenic |
Mag-F
|
Mag-G
|
|
|
Specifications are subject to change without notice.
|